 |
pH
sensor
What is the buffer solution in the little plastic
container that the pH electrode sits in? I need
to know so I can replenish the buffer solution when
my students spill it. |
 |
The solution comprises pH 4 and
1M KCl (in equal amounts)
Back
to TriLog FAQs

|
 |
heart rate
sensor
What settings should I use for the heart rate sensor? |
 |
To read heart rate measurements
TriLog must be connected to the computer and to
work with MultiLab. The sampling rate should be
10/s or higher (for best results use rate of 20/s).
You can also take heart rate measurements disconnected
from the computer. Then download the data to MultiLab.
Back
to TriLog FAQs

|
 |
Dissolved
Oxygen
What settings should I use for the DO2 sensor? How
do I calibrate it? |
 |
We recommend sampling rate of '‘every
second' or less but it will work well with every
rate.
To calibrate the sensor:
Setup TriLog
1. Turn on TriLog
2. Connect Temperature sensor into input 1
(I/O-1) of TriLog
3. Connect the Dissolved Oxygen electrode to
the Dissolved Oxygen adaptor, then plug the
Dissolved Oxygen sensor into input 2 (I/O-2)
of TriLog
4. Set TriLog to record at a rate of every
second
5. Leave the Dissolved Oxygen sensor connected
to TriLog for at least 10 minutes in order to
stabilized
Measure the temperature
Start logging. Wait for stable temperature reading.
Record the temperature
Find the saturation level
Find the correct saturation level for this temperature
in the DO solubility table (see Table 1 below)
Table 1: DO solubility table

Calibrate
1. Push RUN on
TriLog's keypad to begin recording
2. Turn the calibration knob
on the Dissolved Oxygen adaptor until the display
will show the desired saturation level
Back
to TriLog FAQs

|
 |
Oxygen
How do I use the oxygen sensor to measure oxygen
level in the air? |
 |
You need to switch TriLog
to 8 inputs mode and select the sensor type manually.
Oxygen level in the air will be displayed as:
O2 25%
To switch to 8
inputs mode:
With MultiLab:
1. Connect TriLog to
the PC
2. Click Logger
on the menu button, then click Auto
ID
To switch to 8 inputs
click Logger on the menu bar then click Auto ID
a second time to uncheck the Auto ID option.
If the sensor doesn’t support auto ID, you
can still work with it providing that the sensor’s
output is 0 to 5V and linear. To do that you will
have first to define it with MultiLab. See TriLog
User Guide for instructions.
With ImagiProbe:
To activate the Auto ID function, from the Investigation
or Trial form:
1. Tap the Menu icon
to display the ImagiProbe 3.0 menus
2. Tap the Options
menu
3. Select Auto
ID Sensors
4. To disable TriLog's
Auto ID mode tap the Off button.
Back
to TriLog FAQs

|
 |
Turbidity
Do I need to calibrate the turbidity sensor? |
 |
The DT095 ships calibrated.
For more accurate measurements however, we recommend
that you recalibrate every time you perform a
new turbidity experiment.
1. Connect the Turbidity
sensor to input 1 on the TriLog
2. Switch
the TriLog to the 8 sensors mode (Same as question
4)
3. manually select Voltage
0-5V in Input 1
First calibrating
point:
1. Prepare a blank by
rinsing the empty cuvette with distilled water,
then fill it to the marked line with distilled
water
2. Close the cuvette
with the cup. Gently wipe the outside with a
soft, lint-free cloth or tissue
3. Holding the cuvette
by the cup, place it into the slot of the Turbidity
Sensor and close the lid
4. Start recording and
measure the output voltage. This value corresponds
to zero NTU
Second calibrating
point:
1. Obtain the cuvette
containing the Turbidity Standard (100 NTU)and
gently invert it four times in order to mix
in any particles that may have settled to the
bottom
2. NOTE:
Do not shake the standard as this will introduce
tiny air bubbles that will affect turbidity
readings
3. Gently wipe the outside
of the cuvette with a soft, lint-free cloth
or tissue
4. Holding the cuvette
by the cup, place it into the slot of the Turbidity
Sensor and close the lid
5. Start recording and
measure the output voltage. This value corresponds
to 100 NTU
Entering the calibration
values:
1. Click Logger
on the menu bar and then click Define
new sensors
2. Select the Turbidity
sensor* . Enter "0" and "100"
in the real value boxes and
the measured output voltages
in the output voltage boxes
3. Click OK
The Turbidity sensor
is now calibrated and you can switch back to
Auto ID mode and begin recording.
Back
to TriLog FAQs

|